ProductCodes
EAN 5708590358116
ETIM EC010100
GMDN 43600
UNSPSC 42192005
Product group
CT 4000 series
Product
CT 4000 change table, manually foldable, electrically height adjustable
Product description
Pressalit offers a variety of mobile and wall-hung, height-adjustable and fixed height shower and changing tables of high quality. They are designed to provide a stable platform for showering and changing adults and children with reduced mobility.
These tables can be used in schools, institutions, hospitals and private homes and are all designed to be durable, comfortable and easy to clean.
The design and flexibility of the tables ensures optimal working conditions for the carer as well as a high level of comfort and safety for the user.
The CT 4000 series has been designed for public changing facilities and schools.
Description
CT 4000 change table, manually foldable.
Electrically height adjustable 700 mm; from 300 mm -1000 mm, with buttons placed in table front.
Incl. safety rail, body belt and wired remote control.
Excl. mounting kit.
Lying surface 813 x 1829 mm in compact laminate.
Outer measurements 1008 mm x 1829 mm.
Excl. mounting kit.
Mounting instructions
Excl. mounting kit. See Technical datasheet (pdf) for information on mounting kits suitable for various wall types.
Intended use
This product is intended for changing a person who is unable to stand or sit. The person is lying horizontally on the table.
Expected service life is 10 years for institutional use, provided that maintenance is carried out as described in the user manual.
Materials
Powder-laquered stainless steel, lying surface in compact laminate.
Ergonomy
The ergonomically designed safety rail has a practical lower central section, which provides better access and work environment for the carer while transferring, changing and dressing the user.
Easy to clean, smooth contact surfaces which do not collect dirt and bacteria.
Product capacity
Maximum user weight (MUM): 227 kg.
Safe working load (SWL): 232 kg.
When the table is loaded with 227 kg, each mounting bolt supports 37,5 kg
This product has been tested statically with a load of 580 kg as the IEC 60601-1 standard requires testing with safety factor 2.5.
Adjustability
Height adjustable
Mains connection (V)
230 V~
Power consumption (maximum load)
420 W
Power consumption (standby)
2 W
Actuator (motor)
LA40 Careline (Linak) 4000 N, IPX6
Control box
CA63 Careline (Linak), IPX6
Wired remote control
HB30 (Linak) IP66
Maximum user weight (MUM)
227 kg
Safe working load (SWL)
232 kg
Cleaning
CT 4000, 4100, 5000
The product is to be cleaned with hot water or a mild soapy cleaning agent.
Do not use chlorine based, corrosive or abrasive cleaners as it may result in the surface being damaged.
The following cleaning agents have been tested by Pressalit and may be used, as long as they are diluted according to the instructions on the container.
DISINFECTANTS:
Clinimax Difficil-S
Erisan OXY+
Sprit (husholdningssprit)
Suma Bac D10
Virkon
Rodalon Desinfektionsspray
Lysoformin Spezial
CaviCide
Sufanios Premium
Tvättsprit 70%
Accel PREVention Wipes Lingettes
NON-DISINFECTANTS:
Suma Star free D1
Taski 100
Taski 200
Milizid Citro
The product must be wiped down with a soft rag wrung out in clear water immediately after cleaning, as some materials may otherwise be discolored and stained, and the surface coating may deteriorate.
Declaration of conformity
CE UK MDR-MASK-RoHS This product is in compliance with all relevant requirements of:
- Regulation (EU) 2017/745 on medical devices
- Council Directive 2006/42/EC on machinery
- Council Directive 2011/65/EC on RoHS. For the UK:
- Medical Device Regulation (MDR) 2002 on medical devices.
- Supply of Machinery (Safety) Regulations 2008
- Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012.
WEEE
Please enquire of your local Pressalit subsidiary or distributor (in countries where there is no subsidiary) that will arrange for recycling of electronic waste in accordance with the WEEE directive.
Warranty
Pressalit makes a commitment to remedy material and manufacturing defects for a period of three years from the invoicing date.